Unknown

Dataset Information

Commercial smartwatch with pulse oximeter detects short-time hypoxemia as well as standard medical-grade device: Validation study.


ABSTRACT:

Objective

We investigated how a commercially available smartwatch that measures peripheral blood oxygen saturation (SpO2) can detect hypoxemia compared to a medical-grade pulse oximeter.

Methods

We recruited 24 healthy participants. Each participant wore a smartwatch (Apple Watch Series 6) on the left wrist and a pulse oximeter sensor (Masimo Radical-7) on the left middle finger. The participants breathed via a breathing circuit with a three-way non-rebreathing valve in three phases. First, in the 2-minute initial stabilization phase, the participants inhaled the ambient air. Then in the 5-minute desaturation phase, the participants breathed the oxygen-reduced gas mixture (12% O2), which temporarily reduced their blood oxygen saturation. In the final st

SUBMITTER: Rafl J 

PROVIDER: S-EPMC9554125 | biostudies-literature | 2022 Jan-Dec

REPOSITORIES: biostudies-literature