Accuracy of the intermittently scanned continuous glucose monitoring system in critically ill patients: a prospective, multicenter, observational study.


ABSTRACT:

Objective

Continuous glucose monitoring (CGM) has the potential to improve glucose control in the intensive care unit (ICU) setting. We sought to evaluate the accuracy of the intermittently scanned CGM (isCGM) system in critically ill patients.

Research design and methods

Adult patients were consecutively enrolled from three ICUs from August 2020 to January 2021. The performance of FreeStyle Libre Pro was evaluated against the venous blood glucose samples as a reference. Numerical accuracy was examined by the mean absolute relative difference (MARD), the Bland-Altman analysis, and the International Organization for Standardization criteria. Clinical accuracy was assessed by performing the Clarke and consensus error grid analysis.

Results

A total of 122 patients were i

SUBMITTER: Huang W 

PROVIDER: S-EPMC9559122 | biostudies-literature | 2022 Dec