Unknown

Dataset Information

Coronary bioresorbable stents: Non-invasive quantitative evaluation of intra- and juxta-stent plaque composition-A computed tomography longitudinal study.


ABSTRACT:

Purpose

Coronary bioresorbable stents (BRS) do not produce blooming artifacts on computed tomography (CT), in contrast to metallic stents, as they are made of a bioresorbable polymer and are radiolucent. They allow to evaluate the coronary plaque beneath. The low-attenuation plaque (LAP) suggests plaque vulnerability and is CT assessable. The aim of our study was to show the possibility of a non-invasive CT evaluation of the volume and the LAP composition of the intra- and juxta-stent plaque.

Methodology

In our prospective longitudinal study, we recruited 27 consecutive patients (35 BRS stents total; mean age 60 +/- 9 years) with bioresorbable stents for a 256-slice ECG-synchronized CT evaluation at 1- and 12-months post stent implantation. Total plaque volume (mm3), absolut

SUBMITTER: Zdanovich E 

PROVIDER: S-EPMC9560491 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature