ABSTRACT: Purpose
To assess the efficacy of an instrument-integrated OCT (iiOCT)-based distance sensor during robotic vitreoretinal surgery using the Preceyes Surgical System (PSS; Preceyes B.V.).Design
Single-center interventional study.Participants
Patients requiring vitreoretinal surgery.Methods
Five patients were enrolled. Standard preoperative OCT images were obtained. After vitrectomy, a predefined set of actions was performed using the iiOCT-based sensor. Images then were processed to assess the signal-to-noise ratio (SNR) at various angles to the retina and at different distances between the instrument tip and the retinal surface. Preoperative and intraoperative OCT images were compared qualitatively and quantitatively.Main outcomes measures
The feasib