ABSTRACT: Purpose
To compare noncontact acoustic microtapping (AμT) OCT elastography (OCE) with destructive mechanical tests to confirm corneal elastic anisotropy.Design
Ex vivo laboratory study with noncontact AμT-OCE followed by mechanical rheometry and extensometry.Participants
Inflated cornea of whole-globe porcine eyes (n = 9).Methods
A noncontact AμT transducer was used to launch propagating mechanical waves in the cornea that were imaged with phase-sensitive OCT at physiologically relevant controlled pressures. Reconstruction of both Young's modulus (E) and out-of-plane shear modulus (G) in the cornea from experimental data was performed using a nearly incompressible transversely isotropic (NITI) medium material model assuming spatial isotropy of