ABSTRACT: Introduction
Programmable shunt valve settings can sometimes be difficult to assess using classic read-out tools, warranting a skull X-ray.Research question
Can we use available head computed tomography (CT) scans to determine the valve settings, in order to obviate the need for additional skull X-rays?Material and methods
The valve setting of two different programmable shunts (Codman Certas Plus® and Sophysa Polaris®) were assessed by two blinded observers in 24 patients using 65 head CT scans (slice thickness ≤2 mm). Using multi-planar reconstruction (MPR) tools, images were resliced according to the direction of the valve, allowing a direct readout of the valve settings. We validated our CT based method against 32 available skull X-rays.Results