Unknown

Dataset Information

Determination of Carbonyl Compounds in Different Work Environments: Comparison between LC-UV/DAD and LC-MS/MS Detection Methods.


ABSTRACT: There were two analytical methods for the determination of 12 carbonyl compounds (CCs) by using liquid chromatography (LC) coupled with mass spectrometry (MS/MS) and diode array detector (UV/DAD) that were developed and applied to 52 samples that were collected in 10 workplaces. Linearity (0.996 < R2 < 0.999), intra-day repeatability (0.7 < RSD% < 10), and inter-day repeatability (5 < RSD% < 16) were acceptable for both techniques, but the highest sensibility of the MS/MS method allowed us to correctly quantify 98% of the samples (versus 32% by UV/DAD). The comparison of the concentrations that were obtained by quantifying the same sample with both techniques showed good agreement for acetaldehyde and formaldehyde (0.1 < % deviation < 30) but much higher

SUBMITTER: Castellani F 

PROVIDER: S-EPMC9565147 | biostudies-literature | 2022 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Sorry, this publication's infomation has not been loaded in the Indexer, please go directly to PUBMED or Altmetric.

Similar Datasets