Unknown

Dataset Information

Evaluation of Coriolis Micro Air Sampling to Detect Volatile and Semi-Volatile Organic Compounds.


ABSTRACT: There are several analytical procedures available for the monitoring of volatile organic compounds (VOCs) in the air, which differ mainly on sampling procedures. The Coriolis micro air sampler is a tool normally designed for biological air sampling. In this paper, the Coriolis micro bio collector is used to evaluate its ability to sample organic contaminants sampling and detecting them when combined GC-MS. We also compare the use of the Coriolis micro with a standardized sampling method, which is the use of a lung box with a Nalophan® bag. The results show that the Coriolis micro sampling method is suitable for the sampling of organic contaminants. Indeed, the Coriolis micro allows to sample and detect mainly semi-volatile molecules, while the lung box/Nalophan® bags

SUBMITTER: Courtier A 

PROVIDER: S-EPMC9572053 | biostudies-literature | 2022 Sep

REPOSITORIES: biostudies-literature