Biosensors Based on Ion-Sensitive Field-Effect Transistors for HLA and MICA Antibody Detection in Kidney Transplantation.
Ontology highlight
ABSTRACT: This work demonstrates the ability of the Ion-Sensitive Field-Effect Transistor (ISFET)-based immunosensor to detect antibodies against the human leukocyte antigen (HLA) and the major histocompatibility complex class-I-related chain A (MICA). The sensing membrane of the ISFET devices was modified and functionalized using an APTES-GA strategy. Surface properties, including wettability, surface thickness, and surface topology, were assessed in each module of the modification process. The optimal concentrations of HLA and MICA proteins for the immobilization were 10 and 50 μg/mL. The dose-response curve showed a detection range of 1.98-40 µg/mL for anti-HLA and 5.17-40 µg/mL for anti-MICA. The analytical precision (%CV) was found to be 10.69% and 8.92% for anti-HLA and -MICA, respectively. Mo