Unknown

Dataset Information

Validation of a multi-frequency bioelectrical impedance analysis device for the assessment of body composition in older adults with type 2 diabetes.


ABSTRACT:

Background

Aging and type 2 diabetes (T2DM) are associated with an increased risk of sarcopenia. Diagnosis of sarcopenia is commonly done using dual-energy X-ray absorptiometry (DXA) in specialized settings. Another available method for assessing body composition is direct segmental multi-frequency bioelectrical impedance analysis (DSMF-BIA). Here, we examine the accuracy of a DSMF-BIA (InBody-770) for assessing body composition in older adults with T2DM when compared to DXA.

Methods

Eighty-four obese/overweight older adults (49 women, 71 ± 5 years) with T2DM who were recruited for the CEV-65 study and had both DSMF-BIA and DXA assessments at baseline were included. The analysis included Bland-Altman plots and intra class correlation coefficients. Sub-analyses were performed

SUBMITTER: Buch A 

PROVIDER: S-EPMC9584890 | biostudies-literature | 2022 Oct

REPOSITORIES: biostudies-literature