In vitro measurements of ultrafiltration precision in hemofiltration and hemodialysis devices used in infants.


ABSTRACT:

Background

To determine in vitro whether infant hemofiltration and hemodialysis devices can reliably deliver precise ultrafiltration (UF) control.

Methods

We tested the Prismaflex, Aquarius and NIDUS devices which have different circuit types, by in vitro testing with a bag of saline set up as a dummy patient, and monitoring fluid shifts by precise weighing. We looked for differences between the UF rates set and achieved and between the UF result the device displays to the clinician and the true volumes removed, which may lead to clinical errors. We performed short studies at UF settings of zero and 40 ml/h, and with and without simulating poor withdrawal and return lines, and simulated a 4-h treatment session.

Results

The Prismaflex setting vs actual errors and displ