Infant lung function: criteria for selecting tidal flow-volume loops.


ABSTRACT:

Background

Tidal flow-volume (TFV) loops are commonly recorded in infants during sleep, due to the more regular breathing patterns compared to the awake state. Standardised deselection of loops outside pre-specified ranges are based on periods of regular breathing, while criteria and available software for visual evaluation of TFV loops are lacking. We aimed to determine the reliability of standardised criteria for manual selection of infant TFV loops.

Methods

Using a pre-defined set of criteria, three independent raters manually evaluated TFV loops among 57 randomly selected awake healthy 3-month-old infants with available TFV measurements in the Scandinavian Preventing Atopic Dermatitis and ALLergies in children (PreventADALL) study. The TFV loops were sampled using the Eco Medics Exhalyzer D. Criteria for selecting TFV loops included reproducible shape and volume with only one peak in tidal expiratory flow (PTEF), excluding loops with no clear or uneven flow towards PTEF. By intraclass coefficient (ICC), the reliability of agreement between raters was determined for the time to PTEF (t PTEF) to expiratory time (t E) and other TFV loop parameters.

Results

Five infants had unsuccessful tests. Among the remaining 52 infants, the raters selected a median of 25, 26 and 15 loops per test. The ICCs (95% CI) were 0.97 (0.92-0.98) for t PTEF/t E, 0.99 (0.99-1.00) for respiratory rate, 0.98 (0.97-0.99) for tidal volume per kg and 0.98 (0.97-0.99) for expiratory volume, reflecting excellent agreement in all categories.

Conclusion

Manual TFV loop selection using standardised criteria provides a reliable alternative for lung function measures in awake infants with interrupted breathing cycles in a real-life setting.

SUBMITTER: Bains KES 

PROVIDER: S-EPMC9589321 | biostudies-literature | 2022 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Background</h4>Tidal flow-volume (TFV) loops are commonly recorded in infants during sleep, due to the more regular breathing patterns compared to the awake state. Standardised deselection of loops outside pre-specified ranges are based on periods of regular breathing, while criteria and available software for visual evaluation of TFV loops are lacking. We aimed to determine the reliability of standardised criteria for manual selection of infant TFV loops.<h4>Methods</h4>Using a pre-defined  ...[more]

Similar Datasets

| S-EPMC9856825 | biostudies-literature
| S-EPMC9574559 | biostudies-literature
| S-EPMC5889239 | biostudies-literature
| S-EPMC4414869 | biostudies-literature
| S-EPMC6774243 | biostudies-literature
2014-04-24 | GSE52142 | GEO
2012-06-24 | GSE30292 | GEO
| S-EPMC5741669 | biostudies-literature
2014-04-24 | E-GEOD-52142 | biostudies-arrayexpress
2012-06-23 | E-GEOD-30292 | biostudies-arrayexpress