New Faecal Calprotectin Assay by IDS: Validation and Comparison to DiaSorin Method
Ontology highlight
ABSTRACT: Background: The faecal calprotectin (FC) measurement is used for inflammatory bowel disease (IBD) diagnosis and follow-up. The aim of this study was to validate for the first time the new IDS FC extraction device and immunoassay kit, and to compare it with the DiaSorin test in patients with and without IBD. Methods: First, the precision of the IDS assay and its stability were assessed. Then, 379 stool extracts were analysed with the IDS kit on iSYS and compared with a DiaSorin Liaison XL assay. Results: The intra- and inter-assay CVs did not exceed 5%. The stool samples were stable up to 4 weeks at −20 °C. Lot-to-lot comparison showed a good correlation (Lot1 = 1.06 × Lot2 + 0.60; p > 0.05). The Passing and Bablok regression showed no significant deviation from linearity between the two me
SUBMITTER: Castiglione V
PROVIDER: S-EPMC9600005 | biostudies-literature | 2022 Sep
REPOSITORIES: biostudies-literature
ACCESS DATA