Impact of arm position on vertebral bone marrow proton density fat fraction in chemical-shift-encoded magnetic resonance imaging: a preliminary study.
Ontology highlight
ABSTRACT: Arm positions employed during magnetic resonance imaging (MRI) can affect magnetic field distribution, which may result in variability in proton density fat fraction (PDFF) measurements. This study evaluated the effect of arm position on lumbar PDFF measured using chemical-shift-encoded MRI (CSE-MRI). Fifteen healthy volunteers from a single-center underwent lumbar CSE-MRI at two different arm positions (side and elevated) using a single 3T scanner. Scans were performed twice in each position. PDFFs of the L1-L5 vertebrae were independently measured by two readers, and reader measurements were compared by calculating intraclass correlation coefficients (ICC). We compared PDFF measurements from two arm positions and from two consecutive scans using the Wilcoxon test and Bland-Altman