Accuracy, interpretability and usability study of a wireless self-guided fetal heartbeat monitor compared to cardiotocography.
Ontology highlight
ABSTRACT: Fetal Cardiography is usually performed using in-hospital Cardiotocographic (CTG) devices to assess fetal wellbeing. New technologies may permit home-based, self-administered examinations. We compared the accuracy, clinical interpretability, and user experience of a patient-administered, wireless, fetal heartbeat monitor (HBM) designed for home use, to CTG. Initially, participants had paired HBM and CTG examinations performed in the clinic. Women then used the HBM unsupervised and rated the experience. Sixty-three women had paired clinic-based HBM and CTG recordings, providing 6982 fetal heart rate measures for point-to-point comparison from 126 min of continuous recording. The accuracy of the HBM was excellent, with limits of agreement (95%) for mean fetal heart rate (FHR) between 0.72 an