ABSTRACT: Objective
Reflex cryptococcal antigen (CrAg) screening of blood specimens with a CD4 count of <100 cells/µL was performed at 45 South African CD4 laboratories using a lateral flow assay (LFA). Our objective was to evaluate the reliability of routine LFA results through comparative interlaboratory testing.Methods
All CrAg-positive and a selected number of CrAg-negative samples from the CD4 laboratories were retested at paired microbiology laboratories using the same LFA. Samples with discordant results were tested at a reference laboratory, using the LFA (with CrAg titers).Results
During interlaboratory testing, 12,502 samples were retested, with 93 (0.7%) discordant results and a between-laboratory agreement of 99.3% (Cohen's kappa, 0.98). The proportion of retested