Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring).


ABSTRACT:

Background

Nowadays, automated blood pressure (BP) monitoring devices are commonly used by patients as a part of standard medical care for hypertension. The timer trigger was modified into a wireless automated home BP monitoring (HBPM) device to expand its potential use as ambulatory BP monitoring. However, the BP measurement accuracy in this modified device remains unknown.

Objective

We aimed to assess the accuracy of Uright model TD 3127AT, which is an automated HBPM device with a timer trigger modification, following an International Organization for Standardization (ISO) 81060-2:2018 guidelines in the Thai population.

Methods

This cross-sectional study included normotensive and hypertensive Thai participants following the ISO 81060-2:2018 guidelines from August 20

SUBMITTER: Laosanguanek W 