Accuracy of calculating mechanical power of ventilation by one commonly used equation.


ABSTRACT: Gattinoni's equation, [Formula: see text], now commonly used to calculate the mechanical power (MP) of ventilation. However, it calculates only inspiratory MP. In addition, the inclusion of PEEP in Gattinoni's equation raises debate because PEEP does not produce net displacement or contribute to MP. Measuring the area within the pressure-volume loop accurately reflects the MP received in a whole ventilation cycle and the MP thus obtained is not influenced by PEEP. The MP of 25 invasively ventilated patients were calculated by Gattinoni's equation and measured by integration of the areas within the pressure-volume loops of the ventilation cycles. The MP obtained from both methods were compared. The effects of PEEPs on MP were also evaluated. We found that the MP obtained from both methods w

SUBMITTER: Wu SH 

PROVIDER: S-EPMC9637605 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature