ABSTRACT: Objective
To investigate the mechanical properties of 23-, 25-, and 27-gauge vitrectomy vitrectors across 3 different vitrectomy systems to inform surgical techniques.Design
An experimental study that did not involve any human subjects.Methods
Nine vitrectors (3 each of 23-, 25-, and 27 gauge) from Alcon, Dutch Ophthalmic Research Center (DORC), and Bausch & Lomb (B/L) were measured. Measurements were performed using electroforce displacement at the tip and 15 mm from the tip. Five measurements were performed at each location, and fully elastic deformation was ensured.Main outcome measures
The main parameter being measured was the force in grams (gf) necessary to deflect the vitrectors vertically downward by 1 mm, either at the tip of the vitrector or 15 mm