Randomized trial of conventional versus radiofrequency needle transseptal puncture for cryoballoon ablation: the CRYO-LATS trial.


ABSTRACT:

Background

Transseptal puncture to achieve left atrial access is necessary for many cardiac procedures, including atrial fibrillation ablation. More recently, there has been an increasing need for left atrial access using large caliber sheaths, which increases risk of perforation associated with the initial advancement into the left atrium. We compared the effectiveness of a radiofrequency needle-based transseptal system versus conventional needle for transseptal access.

Methods

This prospective controlled trial randomized 161 patients with symptomatic paroxysmal atrial fibrillation undergoing cryoballoon pulmonary vein isolation to transseptal access with a commercially available transseptal system (radiofrequency needle plus stiff pigtail wire; RF + Pigtail group) versus conventional transseptal access (standard group). The primary outcome was time required for left atrial access. Secondary outcomes included failure of the assigned transseptal system, radiation exposure, and complications.

Results

The median transseptal puncture time was significantly shorter using the radiofrequency needle plus stiff pigtail wire transseptal system compared with conventional transseptal (840 ± 323 vs. 956 ± 407 s, P = 0.0489). Compared to conventional transseptal puncture, fewer transseptal attempts were required (1.0 ± 0.5 RF applications vs. 1.3 ± 0.8 mechanical punctures, P = 0.0123) and the fluoroscopy time was significantly shorter (72.0 [IQR 48.0, 129.0] vs. 93.0 [IQR 60.0, 171.0] s, P = 0.0490) with the radiofrequency needle plus stiff pigtail wire transseptal system. Failure to achieve transseptal LA access with the assigned system was rarely observed (1.3% vs. 5.7%, P = 0.2192). There were no procedural complications observed with either system.

Conclusions

The use of a radiofrequency needle plus stiff pigtail wire resulted in shorter time to left atrial access and reduced fluoroscopy time compared to left atrial access using conventional transseptal equipment.

Trial registration

ClinicalTrials.gov identifier NCT03199703.

SUBMITTER: Andrade JG 

PROVIDER: S-EPMC9640463 | biostudies-literature | 2022 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Randomized trial of conventional versus radiofrequency needle transseptal puncture for cryoballoon ablation: the CRYO-LATS trial.

Andrade Jason G JG   Macle Laurent L   Bennett Matthew T MT   Hawkins Nathaniel M NM   Essebag Vidal V   Champagne Jean J   Roux Jean-Francois JF   Makanjee Bhavanesh B   Tang Anthony A   Skanes Allan A   Khaykin Yaariv Y   Morillo Carlos C   Jolly Umjeet U   Lockwood Evan E   Amit Guy G   Angaran Paul P   Sapp John J   Wardell Stephan S   Wells George A GA   Verma Atul A   Deyell Marc W MW  

Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing 20220624 2


<h4>Background</h4>Transseptal puncture to achieve left atrial access is necessary for many cardiac procedures, including atrial fibrillation ablation. More recently, there has been an increasing need for left atrial access using large caliber sheaths, which increases risk of perforation associated with the initial advancement into the left atrium. We compared the effectiveness of a radiofrequency needle-based transseptal system versus conventional needle for transseptal access.<h4>Methods</h4>T  ...[more]

Similar Datasets

| S-EPMC11649652 | biostudies-literature
| S-EPMC12728985 | biostudies-literature
| S-EPMC10794817 | biostudies-literature
| S-EPMC8360464 | biostudies-literature
| S-EPMC12399136 | biostudies-literature
| S-EPMC4473632 | biostudies-literature