Comprehensive clinical evaluation of TomoEQA for patient-specific pre-treatment quality assurance in helical tomotherapy.


ABSTRACT:

Background

Based on a previous study on the feasibility of TomoEQA, an exit detector-based patient-specific pre-treatment quality assurance (QA) method for helical tomotherapy, an in-depth clinical evaluation was conducted.

Methods

Data of one hundred patients were analyzed to evaluate the clinical usefulness of TomoEQA for patient-specific pre-treatment QA in comparison with the conventional phantom-based method. Additional investigations were also performed under unusual measurement conditions to validate the off-axis region. In addition to the clinical evaluation of TomoEQA, a statistical analysis was conducted to determine the plan parameters that affect the pass/failure results of pre-treatment QA.

Results

The average and standard deviations of the gamma passing