Effect of linear accelerator carbon fiber couch on radiotherapy dose.
Ontology highlight
ABSTRACT: This study aimed to explore the effect of carbon fiber couch on radiotherapy dose attenuation and gamma pass rate in intensity-modulated radiotherapy (IMRT) plans. A phantom inserted with an ionization chamber was placed at different positions of the couch, and the dose was measured by the chamber. Under the same positioning, the phantom dose was calculated using the real and virtual couch images, and the difference in the planned dose of radiotherapy was compared. Ten clinical IMRT plans were selected as dose verification data, and the gamma pass rates were compared between couch addition and non-addition conditions. When the radiation field was near 110° and 250°, the measured value attenuation coefficient of the ionization chamber at the joint of the couch was up to 34%; the attenuation