Development of reference intervals for pupillometry in healthy dogs.


ABSTRACT:

Background

Pupillometry, the measurement of pupil size and reactivity to a stimulus, has various uses in both human and veterinary medicine. These reflect autonomic tone, with the potential to assess nociception and emotion. Infrared pupillometry reduces inaccuracies that may occur when the pupillary light reflex is determined subjectively by the examiner. To our knowledge, there are no published studies outlining normal reference intervals for automated pupillometry in dogs.

Objective

The objective of this study was to develop de novo automated pupillometry reference intervals from 126 healthy canine eyes.

Methods

The pupillary light reflex (PLR) was measured with a handheld pupillometer (NeurOptics™ PLR-200™ Pupillometer). Parameters recorded included maximum

SUBMITTER: Mills EP 

PROVIDER: S-EPMC9643214 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature