Validity and reliability of a novel 3D ultrasound approach to assess static lengths and the lengthening behavior of the gastrocnemius medialis muscle and the Achilles tendon in vivo.


ABSTRACT:

Purpose

Human muscle-tendon units (MTUs) are highly plastic and undergo changes in response to specific diseases and disorders. To investigate the pathological changes and the effects of therapeutic treatments, the use of valid and reliable examination methods is of crucial importance. Therefore, in this study, a simple 3D ultrasound approach was developed and evaluated with regard to: (1) its validity in comparison to magnetic resonance imaging (MRI) for the assessment of the gastrocnemius medialis (GM) MTU, muscle belly, and Achilles tendon lengths; and (2) its reliability for static and dynamic length measurements.

Methods

Sixteen participants were included in the study. To evaluate the validity and reliability of the novel 3D ultrasound approach, two ultrasound measureme