Unknown

Dataset Information

Caution on Using Tetrahydrofuran for Processing Crystalline Silica Samples From Engineered Stone for XRD Analysis.


ABSTRACT: We conducted laboratory experiments to investigate a suspected effect of tetrahydrofuran (THF) on quantifying crystalline silica in samples collected from working with engineered stone when THF is used to process samples prior to the X-ray diffraction (XRD) analysis. Two groups of samples from grinding either engineered stone or granite were simultaneously taken from a laboratory testing system, with one group of samples using THF for processing and another group using muffle furnace for ashing. For each stone type, we also tested four levels of respirable dust loading on the samples by varying the grinding time from 1 to 8 min. Statistical analysis of the experimental results on crystalline silica contents of the two groups of samples showed that the difference between the two methods was

SUBMITTER: Qi C 

PROVIDER: S-EPMC9669205 | biostudies-literature | 2022 Nov

REPOSITORIES: biostudies-literature