Pediatric reference interval verification for 17 specialized immunoassays and cancer markers on the Abbott Alinity i system in the CALIPER cohort of healthy children and adolescents.


ABSTRACT:

Objectives

Clinical laboratory investigation of autoimmune, metabolic, and oncologic disorders in children and adolescents relies on appropriateness of reference intervals (RIs). The Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) previously established comprehensive pediatric RIs for specialized immunoassays on the Abbott ARCHITECT system. Herein, we aim to verify performance on new Alinity i assays by evaluating sera collected from healthy children as per Clinical and Laboratory Standards Institute (CLSI) EP-28A3C guidelines.

Methods

Precision, linearity, and method comparison experiments were completed for 17 specialized Alinity immunoassays, including cancer antigens, autoimmune peptides, and hormones. Sera collected from healthy children and ad

SUBMITTER: Bohn MK 

PROVIDER: S-EPMC9691267 | biostudies-literature | 2023 Jan

REPOSITORIES: biostudies-literature