Introduction

Percutaneous fetoscopic surgery is hampered by an increased risk of preterm prelabor rupture of membranes (PPROM). Recent surgical techniques have shown that suturing the chorioamniotic membranes following laparotomy and uterine exteriorization is associated with a lower risk of PPROM compared to percutaneous in utero surgery. This study presents the ChorioAnchor, a novel resorbable device that percutaneously anchors the chorioamniotic membranes to the uterine wall.

Methods

Human factors testing and peel tests were used to simulate the worst-case in-use loading conditions, establishing the device strength requirements. Tensile testing was used to measure the time-zero strength of the device. Porcine cadaver testing was used to examine ultrasound visibility and a