The effect of respiration-induced target motion on 3D magnetic resonance images used to guide radiotherapy.


ABSTRACT:

Background and purpose

3D Magnetic Resonance Imaging (MRI) is used in radiation therapy for reference planning and, lately, for adaptive treatments on MR accelerators. This study aimed to investigate the impact of different types of respiratory motion on the apparent target position and extent in such scans.

Materials and methods

An MRI motion phantom with a 30 mm diameter target was used to simulate cranial-caudal (CC) motion and imaged at an MR-Linac using a standard clinically released 3D T2w sequence. Scans were acquired for each combination of functions (sin(t), sin4(t) and sin12(t)), peak-to-peak amplitudes (5, 10, 15 and 20 mm), and periods (4, 5 and 6 s). Furthermore, respiration CC motion patterns from two patients were used. Motion functions w