The Development of New Methodology for Determination of Vincristine (VCR) in Human Serum Using LC-MS/MS-Based Method for Medical Diagnostics.
Ontology highlight
ABSTRACT: In this article, we have presented the development and validation of a rapid and sensitive reversed-phase liquid chromatography with tandem mass spectrometry (LC-MS/MS) method for the determination of vincristine (VCR) in patient serum samples. Chromatographic separation was achieved on a Kinetex® (Singapore) column using a mobile phase consisting of 25 mM acetic acid and 0.3% formic acid (A) and methanol (B) in a gradient elution mode at a flow rate of 0.3 mL/min. The VCR and internal standard (vinblastine) were monitored using the multiple reaction monitoring mode under positive electrospray ionization. The lower limit of quantification (LLOQ) was 0.67 ng/mL, and the upper limit of quantification (ULOQ) was 250 ng/mL for VCR. The calculated values of LOD and LOQ for VCR were 0
SUBMITTER: Jurowski K
PROVIDER: S-EPMC9694046 | biostudies-literature | 2022 Nov
REPOSITORIES: biostudies-literature
ACCESS DATA