Portable ultrasound assessment of jugular venous pressure is an accurate method for estimating volaemic status in patients with cardiac disease.


ABSTRACT:

Purpose

The objective of this study was to determine whether ultrasound-measured jugular venous pressure (U-JVP) could accurately estimate central venous pressure (CVP).

Methods

This prospective, diagnostic, single-centre study was performed at the Cardiac Intensive Care Unit of the Northern General Hospital, Sheffield, UK. Post-cardiac surgery patients were recruited from January to May 2019. The investigators were blinded to the central venous pressure when measuring the jugular venous pressure. U-JVP and direct CVP were measured simultaneously. Measurements were taken whilst the patient was ventilated and then repeated when the patient was extubated, providing non-ventilated readings.

Results

One-hundred and fourteen consecutive participants with a male predominance of 71% and mean age of 65 ± 12 years were included in the analysis. Bland-Altman plots revealed that U-JVP marginally overestimated CVP by 0.91 mmHg (95% limits of agreement were -2.936 to 4.754) in ventilated patients and by 0.11 mmHg (95% limits of agreement between -2.481 and 2.695) in non-ventilated patients. Reasonable sensitivity and specificity of ultrasound-measured jugular venous pressure was achieved for low and high central venous pressure in both ventilated and non-ventilated patients.

Conclusion

U-JVP accurately estimates cardiac filling pressure and fluid status in patients after cardiac surgery, irrespective of their ventilatory status. Jugular venous pressure measurement by insonation is a reliable technique that can be taught to medical students and other health professionals to non-invasively estimate central venous pressure and may be useful for assessment of volaemic status in patients with heart failure.

Trial registration

ClinicalTrials.gov public (identifier NCT03815188).

SUBMITTER: Jenkins S 

PROVIDER: S-EPMC9705668 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Portable ultrasound assessment of jugular venous pressure is an accurate method for estimating volaemic status in patients with cardiac disease.

Jenkins Sam S   Knowles Patrick P   Briffa Norman N  

Journal of ultrasound 20220315 4


<h4>Purpose</h4>The objective of this study was to determine whether ultrasound-measured jugular venous pressure (U-JVP) could accurately estimate central venous pressure (CVP).<h4>Methods</h4>This prospective, diagnostic, single-centre study was performed at the Cardiac Intensive Care Unit of the Northern General Hospital, Sheffield, UK. Post-cardiac surgery patients were recruited from January to May 2019. The investigators were blinded to the central venous pressure when measuring the jugular  ...[more]

Similar Datasets

| S-EPMC7592775 | biostudies-literature
| S-EPMC9883282 | biostudies-literature
| S-EPMC5628241 | biostudies-literature
| S-EPMC5501709 | biostudies-other
| S-EPMC11607288 | biostudies-literature
| S-EPMC8504525 | biostudies-literature
| S-EPMC9592330 | biostudies-literature
| S-EPMC6930831 | biostudies-literature
| S-EPMC9191400 | biostudies-literature
| S-EPMC7326219 | biostudies-literature