ABSTRACT: Introduction
High frequency ultrasound has shown as a promising imaging modality to evaluate peri-implant tissues. It is not known if the ultrasound imaging settings might influence ultrasound’s ability to differentiate implant structures. The aim of this benchtop study was to evaluate the dependence of ultrasound on imaging angles and modes to measure implant geometry-related parameters. Methods
A clinical ultrasound scanner (ZS3, Mindray) with an intraoral probe (L30-8) offering combinations of harmonic and compound imaging modes was employed for imaging 16 abutments and 4 implants. The samples were mounted to a micro-positioning system in a water tank, which allowed a range of -30 to 30-degree imaging angles in 5-degree increment between the probe and samples. The abut