Automated detection of low ejection fraction from a one-lead electrocardiogram: application of an AI algorithm to an electrocardiogram-enabled Digital Stethoscope.


ABSTRACT:

Aims

Electrocardiogram (ECG)-enabled stethoscope (ECG-Scope) acquires a single-lead ECGs during cardiac auscultation and may facilitate real-time screening for pathologies not routinely identified by cardiac auscultation alone. We previously demonstrated an artificial intelligence (AI) algorithm can identify left ventricular dysfunction (LVSD) [defined as ejection fraction (EF) ≤ 40%] with an area under the curve (AUC) of 0.91 using a 12-lead ECG.

Methods and results

One hundred patients referred for clinically indicated echocardiography were prospectively recruited. ECG-Scope recordings with the patient supine and sitting were obtained in multiple electrode locations at the time of the echocardiogram. The AI algorithm for the detection of LVSD was retrained using single leads from ECG-12 and validated against ECG-Scope to determine accuracy for low EF detection (≤35%, <40%, or <50%). We evaluated the algorithm with respect to body position and lead location. Amongst 100 patients (aged 61.3 ± 13.8; 61% male, BMI: 30.0 ± 5.4), eight had EF≤40%, and six had EF 40-50%. The best single recording position was V2 with the patient supine [AUC: 0.88 (CI: 0.80-0.97) for EF≤35%, 0.85 (CI: 0.75-0.95) for EF≤40%, and 0.81 (CI: 0.71-0.90) for EF < 50%]. When using an AI model to select the recording automatically, AUC was 0.91 (CI: 0.84-0.97) for EF≤35%, 0.89 (CI: 0.83-0.96) for EF≤40%, and 0.84 (CI: 0.73-0.94) for EF < 50%.

Conclusion

An AI algorithm applied to an ECG-enabled stethoscope recording in standard auscultation positions reliably detected the presence of a low EF in this prospective study of patients referred for echocardiography. The ability to screen patients with a possible low EF during routine physical examination may facilitate rapid detection of LVSD.

SUBMITTER: Attia ZI 

PROVIDER: S-EPMC9708035 | biostudies-literature | 2022 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Automated detection of low ejection fraction from a one-lead electrocardiogram: application of an AI algorithm to an electrocardiogram-enabled Digital Stethoscope.

Attia Zachi I ZI   Dugan Jennifer J   Rideout Adam A   Maidens John N JN   Venkatraman Subramaniam S   Guo Ling L   Noseworthy Peter A PA   Pellikka Patricia A PA   Pham Steve L SL   Kapa Suraj S   Friedman Paul A PA   Lopez-Jimenez Francisco F  

European heart journal. Digital health 20220523 3


<h4>Aims</h4>Electrocardiogram (ECG)-enabled stethoscope (ECG-Scope) acquires a single-lead ECGs during cardiac auscultation and may facilitate real-time screening for pathologies not routinely identified by cardiac auscultation alone. We previously demonstrated an artificial intelligence (AI) algorithm can identify left ventricular dysfunction (LVSD) [defined as ejection fraction (EF) ≤ 40%] with an area under the curve (AUC) of 0.91 using a 12-lead ECG.<h4>Methods and results</h4>One hundred p  ...[more]