ABSTRACT: Primary objective
To evaluate the effect of latex tip cover manufacturer on accuracy and repeatability of Tono-Pen Vet™ in canine eyes.Animal studied
Twelve enucleated globes from six dogs.Procedures
The anterior chamber was cannulated and connected to a calibrated manometer. Intraocular pressure (IOP) measurements were obtained using the Tono-Pen Vet and TONOVET Plus at manometric IOP ranging from 5 to 80 mmHg. At each IOP, the Tono-Pen Vet was used with a new Ocu-Film™ latex tip cover (the only manufacturer-approved brand of cover) followed by a new Softips™ latex tip cover. For comparison, the TONOVET Plus was also used at each IOP with a new disposable rebound probe. Measured IOP values were analyzed by linear regression and intraclass correlation coefficient (I