Comparison of dosimetric effects of MLC positional errors on VMAT and IMRT plans for SBRT radiotherapy in non-small cell lung cancer.
Ontology highlight
ABSTRACT: The positional accuracy of multi-leaf collimators (MLC) is important in stereotactic body radiotherapy (SBRT). The aim of this study was to investigate the impact between MLC positional error and dosimetry of volume intensity modulated (VMAT) and general intensity modulated (IMRT) plans for non-small cell lung cancer (NSCLC). Fifteen patients with NSCLC were selected to design the 360 SBRT-VMAT plans and the 360 SBRT-IMRT error plans. The DICOM files for these treatment plans were imported into a proprietary computer program that introduced delivery errors. Random and systematic MLC position (0.1, 0.2, 0.5, 1.0, 1.5, and 2.0 mm) errors were introduced. The systematic errors were shift errors (caused by gravity), opening errors, and closing errors. The CI, GI, d2cm and generalized equivalen