Impact of reference point selection on DXA-based measurement of forearm bone mineral density.


ABSTRACT: Few studies have systematically evaluated the technical aspects of forearm bone mineral density (BMD) measurement. We found that BMD remained stable regardless of the reference point; however, the ROI identified was not always consistent. Our study highlights the importance of using the same reference point for serial measurements of forearm BMD.

Background

Forearm fractures are clinically important outcomes from the perspective of morbidity, health care costs, and interruption of work. BMD of the forearm, as derived by dual-energy x-ray absorptiometry (DXA), is a better predictor of fracture at the forearm compared with BMD measured at other sites. However, very few studies have evaluated the technical aspects of selecting the ROI for forearm BMD measurement. This study aimed to co