Chronic haemodynamic performance of a biorestorative transcatheter heart valve in an ovine model.


ABSTRACT:

Background

The Xeltis biorestorative transcatheter heart valve (BTHV) leaflets are made from an electrospun bioabsorbable supramolecular polycarbonate-urethane and are mounted on a self-expanding nitinol frame. The acute haemodynamic performance of this BTHV was favourable.

Aims

We sought to demonstrate the preclinical feasibility of a novel BTHV by evaluating the haemodynamic performances of five pilot valve designs up to 12 months in a chronic ovine model.

Methods

Five design iterations (A, B, B', C, and D) of the BTHV were transapically implanted in 46 sheep; chronic data were available in 39 animals. Assessments were performed at implantation, 3, 6, and 12 months including quantitative aortography, echocardiography, and histology.

Results

At 12 months, grea

SUBMITTER: Serruys PW 

PROVIDER: S-EPMC9725010 | biostudies-literature | 2021 Dec