High-Throughput Steady-State Enzyme Kinetics Measured in a Parallel Droplet Generation and Absorbance Detection Platform.
Ontology highlight
ABSTRACT: Microfluidic water-in-oil emulsion droplets are becoming a mainstay of experimental biology, where they replace the classical test tube. In most applications, such as ultrahigh-throughput directed evolution, the droplet content is identical for all compartmentalized assay reactions. When emulsion droplets are used for kinetics or other functional assays, though, concentration dependencies of initial rates that define Michaelis-Menten parameters are required. Droplet-on-demand systems satisfy this need, but extracting large amounts of data is challenging. Here, we introduce a multiplexed droplet absorbance detector, which─coupled to semi-automated droplet generation─forms a tubing-based droplet-on-demand system able to generate and extract quantitative datasets from defined concentration gr
SUBMITTER: Neun S
PROVIDER: S-EPMC9730296 | biostudies-literature | 2022 Dec
REPOSITORIES: biostudies-literature