Biomechanical evaluation of aortic regurgitation from cusp prolapse using an ex vivo 3D-printed commissure geometric alignment device.


ABSTRACT:

Background

Aortic regurgitation (AR) is one of the most common cardiac valvular diseases, and it is frequently caused by cusp prolapse. However, the precise relationship of commissure position and aortic cusp prolapse with AR is not fully understood. In this study, we developed a 3D-printed commissure geometric alignment device to investigate the effect of commissure height and inter-commissure angle on AR and aortic cusp prolapse.

Methods

Three porcine aortic valves were explanted from hearts obtained from a meat abattoir and were mounted in the commissure geometric alignment device. Nine commissure configurations were tested for each specimen, exploring independent and concurrent effects of commissure height and inter-commissure angle change on AR and aortic cusp prolapse. Each commissure configuration was tested in our 3D printed ex vivo left heart simulator. Hemodynamics data, echocardiography, and high-speed videography were obtained.

Results

AR due to aortic cusp prolapse was successfully generated using our commissure geometric alignment device. Mean aortic regurgitation fraction measured for the baseline, high commissure, low commissure, high commissure and wide inter-commissure angle, high commissure and narrow inter-commissure angle, low commissure and wide inter-commissure angle, low commissure and narrow inter-commissure angle, wide commissure, and narrow commissure configurations from all samples were 4.6 ± 1.4%, 9.7 ± 3.7%, 4.2 ± 0.5%, 11.7 ± 5.8%, 13.0 ± 8.5%, 4.8 ± 0.9%, 7.3 ± 1.7%, 5.1 ± 1.2%, and 7.1 ± 3.1%, respectively.

Conclusions

AR was most prominent when commissure heights were changed from their native levels with concomitant reduced inter-commissure angle. Findings from this study provide important evidence demonstrating the relationship between commissure position and aortic cusp prolapse and may have a significant impact on patient outcomes after surgical repair of aortic valves.

SUBMITTER: Zhu Y 

PROVIDER: S-EPMC9737730 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Biomechanical evaluation of aortic regurgitation from cusp prolapse using an ex vivo 3D-printed commissure geometric alignment device.

Zhu Yuanjia Y   Park Matthew H MH   Imbrie-Moore Annabel A   Wilkerson Robert R   Madira Sarah S   Woo Y Joseph YJ  

Journal of cardiothoracic surgery 20221210 1


<h4>Background</h4>Aortic regurgitation (AR) is one of the most common cardiac valvular diseases, and it is frequently caused by cusp prolapse. However, the precise relationship of commissure position and aortic cusp prolapse with AR is not fully understood. In this study, we developed a 3D-printed commissure geometric alignment device to investigate the effect of commissure height and inter-commissure angle on AR and aortic cusp prolapse.<h4>Methods</h4>Three porcine aortic valves were explante  ...[more]