Mapping Pulsatile Optic Nerve Head Deformation Using OCT.


ABSTRACT:

Objective

To develop a noninvasive technique to quantitatively assess the pulsatile deformation due to cardiac contractions of the optic nerve head (ONH).

Design

Evaluation of a diagnostic test or technology.

Participants

Healthy subjects with no history of refractive surgery, divided into 2 cohorts on the basis of their axial length (AL).

Methods

We present a noninvasive technique to quantitatively assess the pulsatile deformation of the ONH tissue by combining high-frequency OCT imaging and widely available image processing algorithms. We performed a thorough validation of the approach, numerically and experimentally, evaluating the sensitivity of the method to artificially induced deformation and its robustness to different noise levels. We performed deformation measurements in cohorts of healthy (n = 9) and myopic (n = 5) subjects in different physiological strain conditions by calculating the amplitude of tissue displacement in both the primary position and abduction. The head rotation was measured using a goniometer. During imaging in abduction, the head was rotated 40° ± 3°, and subjects were instructed to direct their gaze toward the OCT visual target.

Main outcome measures

Pulsatile tissue displacement maps.

Results

The robustness of the method was assessed using artificial deformations and increasing noise levels. The results show acceptable absolute errors before the noise simulations grossly exaggerate image degradation. For the group of subjects with AL of < 25 mm (n = 9), the median pulsatile displacement of the ONH was 7.8 ± 1.3 μm in the primary position and 8.9 ± 1.2 μm in abduction. The Wilcoxon test showed a significant difference (P ≤ 0.005) between the 2 paired measures. Reproducibility was tested in 2 different sessions in 5 different subjects with the same intraocular pressure, and an intraclass correlation coefficient of 0.99 was obtained (P < 0.005).

Conclusions

The computational pipeline demonstrated good reproducibility and had the capacity to accurately map the pulsatile deformation of the optic nerve. In a clinical setting, we detected physiological changes in normal subjects supporting its translation potential as a novel biomarker for the diagnosis and progression of optic nerve diseases.

SUBMITTER: Solano MM 

PROVIDER: S-EPMC9754981 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Mapping Pulsatile Optic Nerve Head Deformation Using OCT.

Solano Marissé Masís MM   Richer Emmanuelle E   Cheriet Farida F   Lesk Mark R MR   Costantino Santiago S  

Ophthalmology science 20220806 4


<h4>Objective</h4>To develop a noninvasive technique to quantitatively assess the pulsatile deformation due to cardiac contractions of the optic nerve head (ONH).<h4>Design</h4>Evaluation of a diagnostic test or technology.<h4>Participants</h4>Healthy subjects with no history of refractive surgery, divided into 2 cohorts on the basis of their axial length (AL).<h4>Methods</h4>We present a noninvasive technique to quantitatively assess the pulsatile deformation of the ONH tissue by combining high  ...[more]

Similar Datasets

| S-EPMC5295769 | biostudies-literature
| S-EPMC6911619 | biostudies-literature
| S-EPMC3017237 | biostudies-literature
| S-EPMC6717668 | biostudies-literature
2013-07-15 | E-GEOD-40857 | biostudies-arrayexpress
| S-EPMC6851461 | biostudies-literature
| S-EPMC8234108 | biostudies-literature
2013-07-15 | GSE40857 | GEO
| S-EPMC8063221 | biostudies-literature
| S-EPMC6312218 | biostudies-literature