Air entrapment as a potential cause of early subcutaneous implantable cardioverter defibrillator malfunction: a systematic review of the literature.


ABSTRACT:

Aims

Air entrapment (AE) has been reported as a potential cause of early inappropriate shocks (ISs) following subcutaneous implantable cardioverter defibrillator (S-ICD) implantation, but a cause-effect relationship is not always evident. This systematic review aims to analyse this phenomenon concerning implantation techniques, electrogram (EGM) features, radiologic findings, and patient management.

Methods and results

A systematic search was conducted using PubMed, Embase, and Google Scholar databases following the PRISMA guidelines to obtain all available literature data since 2010 on S-ICD malfunctions possibly due to AE. The final analysis included 54 patients with AE as a potential cause of S-ICD malfunction. Overall, the aggregate incidence of this condition was 1.2%. Of ICD malfunctions possibly due to AE, 93% were ISs, and 95% were recorded within the first week following implantation. Radiologic diagnosis of AE was confirmed in 28% of the entire study cohort and in 68% of patients in whom this diagnostic examination was reported. At the time of device malfunction, EGMs showed artefacts, baseline drift, and QRS voltage reduction in 95, 76, and 67% of episodes, respectively. Management included ICD reprogramming or testing, no action (observation), and invasive implant revision in 57, 33, and 10% of patients, respectively. No recurrences occurred during follow-up, irrespective of management performed.

Conclusions

Device malfunction possibly due to AE may occur in ∼1% of S-ICD recipients. Diagnosis is strongly suggested by early occurrence, characteristic EGM features, and radiologic findings. Non-invasive management, principally device reprogramming, appears to be effective in most patients.

SUBMITTER: Ali H 

PROVIDER: S-EPMC9757921 | biostudies-literature | 2022 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Air entrapment as a potential cause of early subcutaneous implantable cardioverter defibrillator malfunction: a systematic review of the literature.

Ali Hussam H   Lupo Pierpaolo P   Foresti Sara S   De Ambroggi Guido G   De Lucia Carmine C   Penela Diego D   Turturiello Dario D   Paganini Edoardo Maria EM   Cappato Riccardo R  

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology 20221001 10


<h4>Aims</h4>Air entrapment (AE) has been reported as a potential cause of early inappropriate shocks (ISs) following subcutaneous implantable cardioverter defibrillator (S-ICD) implantation, but a cause-effect relationship is not always evident. This systematic review aims to analyse this phenomenon concerning implantation techniques, electrogram (EGM) features, radiologic findings, and patient management.<h4>Methods and results</h4>A systematic search was conducted using PubMed, Embase, and Go  ...[more]

Similar Datasets

| S-EPMC10455445 | biostudies-literature
| S-EPMC8356671 | biostudies-literature
| S-EPMC7197388 | biostudies-literature
| S-EPMC9544305 | biostudies-literature
| S-EPMC4802466 | biostudies-other
| S-EPMC6630194 | biostudies-literature
| S-EPMC5419807 | biostudies-literature
| S-EPMC7519585 | biostudies-literature
| S-EPMC7607386 | biostudies-literature