Tracheal airway pressure in tracheostomy tube capping trials: an experimental study.


ABSTRACT:

Background

Tracheostomy tube capping is a commonly used test to determine if the tracheostomy tube can be removed. The success of the capping trial depends on the patient's ability to maintain sufficient spontaneous breathing with an occluded tracheostomy tube. The impact of an occluded tracheotomy tube on airway resistance is currently unknown. The aim of this study was to investigate tracheal pressure during capping or stoma button insertion and potential determinants concerning cuff.

Methods

Eight cuffed and uncuffed tracheostomy tubes and three stoma buttons of various manufacturers and sizes were inserted into the trachea model. Cuffs were completely deflated or contained atmospheric pressure. The trachea was ventilated bidirectional with a respirator in volume-controll

SUBMITTER: Nowak A 

PROVIDER: S-EPMC9768925 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature