Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits.


ABSTRACT:

Objective

Acute tissue ischaemia may arise due to arterial emergencies or during more complex vascular procedures and may be mitigated by temporary shunting techniques.
Endovascular shunting (ES) techniques enable percutaneous access and shunting from the donor artery without the need to completely interrupt the arterial flow in the donor artery. An endoshunt system may also cover longer distances than most conventional shunts. The aim was to investigate and optimise the flow rates in different endovascular shunt systems.

Methods

Step 1: The flow capacity of different ES configurations was compared with the flow capacity of a 9 Fr Pruitt-Inahara shunt (PIS). An intravenous bag with 0.9% NaCl, pressurised to 90 mmHg, was connected simultaneously to a PIS and to one of the tes