Influence of Graft Positioning during the Latarjet Procedure on Shoulder Stability and Articular Contact Pressure: Computational Analysis of the Bone Block Effect.
Ontology highlight
ABSTRACT: The Latarjet procedure is the most popular surgical procedure to treat anterior glenohumeral (GH) instability in the presence of large anterior glenoid bone defects. Even though the placement of the bone graft has a considerable influence on its efficacy, no clear indications exist for the best graft position. The aim of this study was to investigate the influence of the medial-lateral positioning of the bone graft on the contact mechanics and GH stability due to the bone block effect. Four finite element (FE) models of a GH joint, with a 20% glenoid bone defect, treated by the Latarjet procedure were developed. The FE models differed in the medial-lateral positioning of the bone graft, ranging from a flush position to a 4.5 mm lateral position with respect to the flush position. All graft