Determination of the Assigned Values of Blood Cells by an Impedance Method for Hematological Reference Samples Used in Hematology External Quality Assessment (EQA) Programs.
Ontology highlight
ABSTRACT: The research into and production of hematological reference samples used to implement an external quality assessment (EQA) to check the quality of hematology tests are necessary for hematology laboratories in Vietnam. In this research, the study team determined the assessment values of blood cell count (human RBCs, pseudo-leucocytes, and pseudo-platelets) by the impedance method used in hematology EQA programs. The hematological reference samples were controlled at three concentration levels: low, normal, and high. Determination of the assigned value (mean ± 2SD) was performed for the following hematology analyzer series by impedance method: ABX Micros 60, Celldyn 1700, and Mindray BC 2000. Each device was sent to 10 different laboratories for evaluation. Research results for assigned valu
SUBMITTER: Vo NN
PROVIDER: S-EPMC9775971 | biostudies-literature | 2022 Dec
REPOSITORIES: biostudies-literature
ACCESS DATA