Anchoring Mechanism for Capsule Endoscope: Mechanical Design, Fabrication and Experimental Evaluation.
Ontology highlight
ABSTRACT: Capsule endoscopes are widely used to diagnose gut-related problems, but they are passive in nature and cannot actively move inside the gut. This paper details the design process and development of an anchoring mechanism and actuation system to hold a capsule in place within the small intestine. The design centres around the mechanical structure of the anchor that makes use of compliant Sarrus linkage legs, which extend to make contact with the intestine, holding the capsule in place. Three variants with 2 legs, 3 legs and 4 legs of the anchoring mechanism were tested using a shape memory alloy spring actuator (5 mm × ϕ 3.4 mm). The experiments determine that all the variants can anchor at the target site and resist peristaltic forces of 346 mN. The proposed design is well suited for an in
SUBMITTER: Rehan M
PROVIDER: S-EPMC9782074 | biostudies-literature | 2022 Nov
REPOSITORIES: biostudies-literature
ACCESS DATA