Unknown

Dataset Information

In Silico Modelling to Assess the Electrical and Thermal Disturbance Provoked by a Metal Intracoronary Stent during Epicardial Pulsed Electric Field Ablation.


ABSTRACT: Background: Pulsed Electric Field (PEF) ablation has been recently proposed to ablate cardiac ganglionic plexi (GP) aimed to treat atrial fibrillation. The effect of metal intracoronary stents in the vicinity of the ablation electrode has not been yet assessed. Methods: A 2D numerical model was developed accounting for the different tissues involved in PEF ablation with an irrigated ablation device. A coronary artery (with and without a metal intracoronary stent) was considered near the ablation source (0.25 and 1 mm separation). The 1000 V/cm threshold was used to estimate the ‘PEF-zone’. Results: The presence of the coronary artery (with or without stent) distorts the E-field distribution, creating hot spots (higher E-field values) in the front and rear of the artery, and cold spots (low

SUBMITTER: Gonzalez-Suarez A 

PROVIDER: S-EPMC9784210 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature