Drops of Capillary Blood Are Not Appropriate for Hemoglobin Measurement with Point-of-Care Devices: A Comparative Study Using Drop Capillary, Pooled Capillary, and Venous Blood Samples.


ABSTRACT: Population-based surveys matched by time but using different methodologies for determining hemoglobin (Hb) concentration have shown inconsistencies in estimating anemia prevalence. This study aimed to estimate measurement errors in Hb quantification in HemoCue 201+ using venous blood (VB) and capillary blood both drops (DCB) and pools (PCB), and compare the results against those of a reference method (VB analyzed in hematology analyzers based on the cyanmethemoglobin method). Children (n = 49), adult females (n = 50), and older adults (n = 50) were randomly allocated to donate VB (4 mL) and either DCB (three drops) or PCB (350 µL). Results in HemoCue were analyzed through Bland Altman and Lyn's concordance against Hb concentration by the reference method. A positive av