Solid Sampling Pyrolysis Adsorption-Desorption Thermal Conductivity Method for Rapid and Simultaneous Detection of N and S in Seafood.
Ontology highlight
ABSTRACT: In this work, a rapid method for the simultaneous determination of N and S in seafood was established based on a solid sampling absorption-desorption system coupled with a thermal conductivity detector. This setup mainly includes a solid sampling system, a gas line unit for controlling high-purity oxygen and helium, a combustion and reduction furnace, a purification column system for moisture, halogen, SO2, and CO2, and a thermal conductivity detector. After two stages of purging with 20 s of He sweeping (250 mL/min), N2 residue in the sample-containing chamber can be reduced to <0.01% to improve the device's analytical sensitivity and precision. Herein, 100 s of heating at 900 °C was chosen as the optimized decomposition condition. After the generated S
SUBMITTER: Shang D
PROVIDER: S-EPMC9786581 | biostudies-literature | 2022 Dec
REPOSITORIES: biostudies-literature
ACCESS DATA