Rapid quantitation of SARS-CoV-2 antibodies in clinical samples with an electrochemical sensor.


ABSTRACT: The current coronavirus disease 2019 (COVID-19) pandemic is caused by several variants of severe acute respiratory syndrome coronavirus-2 virus (SARS-CoV-2). With the roll-out of vaccines and development of new therapeutics that may be targeted to distinct viral molecules, there is a need to screen populations for viral antigen-specific SARS-CoV-2 antibodies. Here, we report a rapid, multiplexed, electrochemical (EC) device with on-chip control that enables detection of SARS-CoV-2 antibodies in less than 10 min using 1.5 μL of a patient sample. The EC biosensor demonstrated 100% sensitivity and specificity, and an area under the receiver operating characteristic curve of 1, when evaluated using 93 clinical samples, including plasma and dried blood spot samples from 54 SARS-CoV-2 positive a