ABSTRACT: Purpose
To evaluate the maximal load at failure, cyclic displacement, and stiffness of onlay subpectoral biceps tenodesis (BT) with an intramedullary unicortical metal button (MB) versus a unicortical all-suture button (ASB).Methods
Eighteen matched paired human cadaveric proximal humeri were randomly allocated for subpectoral BT with either ASB or MB using a high-strength suture. Specimens were tested on a servohydraulic mechanical testing apparatus under cyclic load for 1,000 cycles and then loaded to failure. The clamp was then adjusted to isolate the suture-anchor point interface and loaded to failure. Maximal load to failure, displacement, and stiffness were compared.Results
There was no significant difference between groups in stiffness, displacement, or yield